Equipment

<< Click to Display Table of Contents >>

Navigation:  Instruments > V-M.O.L.E.® >

Equipment

The V-M.O.L.E.® Profiler is a compact and durable thermal profiling instrument ECD offers. One end has input connectors for attaching up to (3) sensors and the other end is the internal Power Pack.

v-mole_reverse

Features/functions:

Thermocouple/Inputs:This is where Type “K” Thermocouple sensors are connected.

Buttons: ON/OFF, Record & OK.

Button:

Action:

ON/OFF

Turns Profiler "ON/OFF".

Record

Starts/Stops Profiler recording data.

OK

Invokes "OK" process where the last recorded profile is compared to pre-configured criteria resulting in a "Pass" or "Fail" mode.

 

Activity indicators: These are LED's that indicate what state the Profiler is in. Refer to the illustration and table below.

LED:

Action:

LED Color

v-mole_top

ON/OFF

Indicates Profiler is "ON" and idle

Green (Solid)

Record

Indicates Profiler is recording data

Green (Flashing)

OK

Indicates recorded profile passes pre-configured criteria using the Target 10-OK tab in the software.

Green - Pass (Solid)

Red - Fail (Solid)

Temp(erature)

Indicates if the internal temperature is at or above a certain threshold.

Red (Solid) >40°C

Battery

Indicates when the internal Power Pack voltage is low

Red (Solid) <3.0V

Charge

Indicates when the internal Power Pack is charging

Yellow (Solid)